Quick Reference Guide

SCRAM Remote Breath Pro
Quick Reference Guide
8
3. Select the serial number hyperlink.
4. Select the Initiate Calibration button.
5. Wake up/power up the device and follow all on screen
prompts.
6. Verify the calibration and calibration check results appear
on the Device Details page in SCRAMNET for the device.
Perform a Calibration Check
1. Ensure the calibration equipment is in good working order
and the canister has an adequate gas level to complete
the calibration check.
2. In SCRAMNET, select the Inventory button and search
for the device's serial number using the lters.
3. Select the serial number hyperlink.
4. Select the Calibration Check button.
5. Wake up/power up the device and follow all on screen
prompts.
6. Verify the calibration check results appear on the Device
Details page in SCRAMNET for the device.
Perform Air-Blank Test
1. In SCRAMNET, select the Inventory button and search
for the device's serial number using the lters.
2. Select the serial number hyperlink.
3. Select the Initiate Air-Blank button.
4. Wake up the device. No further action is required on
your part. The device will step through the air-blank test
process itself.
5. Verify the Air-Blank test results appear on the Device
Details page in SCRAMNET for the device.