Quick Reference Guide

SCRAM Remote Breath Pro
Quick Reference Guide
7
Device Calibration
To ensure accurate readings, the SCRAM Remote Breath
Pro breath-alcohol testing device must be calibrated at least
once every 12 months. Alcohol Monitoring Systems, Inc. (d/b/a
“SCRAM Systems”) will not endorse the breath alcohol test
results of a device that was not calibrated at least 12 months
prior to the date of a given test result.
Calibrating the SCRAM Remote Breath Pro (RB Pro) device
is a simple process that takes just a few minutes to complete.
Components needed to calibrate a device include:
• Gas Canister
• Regulator
• Rubber Tubing
• RB Pro Breath Tube
Calibrations and calibration checks must be performed by
an AMS-certied calibration technician. To become certied,
send an e-mail to training@scramsystems.com requesting
enrollment in the SCRAM Remote Breath Calibration
Certication Training course.

Scan the QR code or visit https://help.scramnetwork.com/
documentation/remote-breath/equipment-management/ for
detailed instructions. (SCRAMNET login required)
1. Ensure the calibration equipment is in good working order
and the canister has an adequate gas level to complete
the calibration and a calibration check.
2. In SCRAMNET, select the Inventory button and search
for the device's serial number using the lters.