Quick Reference Guide

SCRAM CAM
Quick Reference Guide
26
7. Verify that re-initialization has started.
a. In SCRAM
NET, access the client’s Reports page.
b. Click the Communication sub-tab.
c. Verify by looking at the top Date/Time link that the base
station has recently communicated with SCRAM
NET.
d. Open the communication and verify that a “Initial Baseline
Reading Taken” message appears.
e. If the message does NOT appear, contact AMS Customer
Services.
8. Insert the tamper clips into the buckle slots and squeeze together
until you hear a distinct click.
PERFORM OBSERVED IR READINGS
1. Wake up the bracelet.
a. Momentarily place the magnet on the top right side of the
bracelet.
In approximately 15 seconds, the bracelet pump will run—
indicated by a buzz.
2. Repeat Step 1a.
a. Repeat step 1a every 60 seconds until the desired number of
readings are taken. AMS recommends at least ve readings.
3. Log into the Direct Connect application.
a. Open the Direct Connect application.
b. Log into the Direct Connect application using your SCRAM
NET
username and password.
4. Connect the Direct Connect device with your computer.
a. Plug the small (mini-B type) USB connector into the Direct
Connect device.
b. Plug the large (A type) USB connector into a USB port on
your computer.
The
red and blue lights on the Direct Connect device blink
for approximately 5 seconds then remain on.
Bracelet status in the Direct Connect application: Waiting for
Bracelet
5. Attach the Direct Connect device to the bracelet.
a. Align the Direct Connect device with the top of the bracelet.
b. Slide the Direct Connect device down over the bracelet.
6. Wake up the bracelet.
a. Momentarily place the magnet on the Direct Connect device
where the word “magnet” appears.
At this time:
» Bracelet status in the Direct Connect application:
Communicating