User's Guide

SCRAM Rem ot e Br eat h
Most exible opt ion in breat h alcohol t esting!
16
CLOSI N G TASKS
CHECK I N SCRAM REM OTE BREATH DEVI CE
Com plet e t he steps below when the client arrives with
the device. I f t he device is NOT on hand, click t he St op
Monit oring but ton, select t he Stop m onit oring now
and check in equipm ent later opt ion, and com plet e t he
St op Monit oring wizard to set t he status to Await ing
Ret urn. The device will st ay in t his st at us unt il:
The device com m unicates with SCRAM NET and
is checked in by clicking t he M ove t o I nven-
t ory link on the client ’s Equipm ent page
You m anually set t o the device t o t he Lost
stat us
30 days elapses
Access t he client s 1. Equipm ent page.
Click the 2. St op Monit oring but ton t o access the
St op Monit oring Wizard page.
Select t he 3. Check in Equipm ent now opt ion and
click t he N ex t but ton.
Press and hold t he 4. Pow er but ton on the device for
approxim ately ve seconds unt il t he bat tery light
turns
gre en.
The device displays PLEASE WAI T, com m unicates
with SCRAMNET, and t ransm it s any rem aining dat a.
The device t hen changes to SHUTTI NG OFF and
turns off.
Click the 5. Finish but ton t o close t he wizard and
ret urn t o the client s Equipm ent page.
The checked- in device is no longer displayed.