Alcatel - High Vacuum Technology - User’s Manual ADP/ADS
Connection to the cooling circuit
1/2
Edition 03 - May 97
Water characteristics
Standard version
connections
In order to limit corrosion and fouling of the motor cooling coils,
it is recommended to use cooling water with the following
characteristics :
Treated fresh water or non-aggressive industrial water
pH between 7.5 and 11
Hardness < 7 milli-equivalent/dm
3
Resistivity > 1500 .cm
Solid pollution < 100 mg/dm
3
Temp. from 10° to 25°C
Pressure range between 3 and 7 bars absolutes
Pressure inlet/outlet 3 bars mimimum
If the above mentioned characteristics are not respected, install
a filter to the cooling connection.
OUT
IN
REAR
VIEW
1/4 NPT
1/4 Gas
1/4 Gas
1/4 NPT
ALCATEL supply