Specifications

Appendix
23.4 Simultaneous operation of input and output
Via the 110 plug you can either
give set values (voltage or current input),
monitor actual values (voltage or current output) or
simultaneously give set values and monitor actual
values. In this case, input and output channels can be
combined at will
(e.g. channel
0
for input and channel
7
for output).
CONFII-OIANALOGIINPUT
23.45
OC
INPUT DEACTIVATED
!
/
CONFII-OIANALOGIOUTPUT
1
23.5 Stopping the input via the I10 plug
To stop the operation of the analog 110 plug, select STOP in
the INPUT or OUTPUT menu, respectively, and confirm with
ENTER. The message "INPUT DEACTIVATED
!"
or "OUT-
PUT DEACTIVATED
!"
is shown.
23.45
OC
OUTPUT DEACTIVATED
!
23.6 Starting by means of the timer (only F8 and N8)
The operation of the unit with the I10 plug can be started by
means of the timer that is built into circulators of the
F8
and
N8 line.
To do so the timer has to be set first. Before reaching the start
time of the timer the parameters of the
I10 plug
can be set.
At the start time of the timer, the unit is started as are the input
or output of data via the
I10 plug.
20.00"C
ALARM
+
T-TSET <5.OC:F01
23.7 Fault display F01
If the circulator switches off this can cause a heavily dis-
turbed voltage or current signal. Please ensure that the devi-
ation of voltage or current is smaller than the equivalent of
0,1
"C.