Specifications
Disassembly
21.
Disassembly
of
Temperature Control
Module
and
Bath Vessel
1
Switch off the unit
and
pull out the mains plug.
Loosen the
2
crosshead screws right beside the
00
Oo*
.
pump connections.
Slide the metal jacket to the front (slightly bending
may be necessary) and remove
it.
Unscrew the
4
screws.
A
hexagon socket
(10
mm) should be
used
if the screws are too tightly
tightened.
In case of difficulty loosen the sealing from the
unit with a screw driver.
Incline the unit slightly whilst lifting it.
!
Lift
the float to avoid any damage.
-
@
Fix the jacket again with the
2