Specifications

Serial
Interface
2.
Then define the
number
of
segments
the program
should have and the
number of program
cycles.
Max.
6
segments can be defined for one program.
A program can be repeated up to 90 times
(i.e.
90
pro-
gram cycles):
W,SP,Q),<number
of
segment~>~0,<nurnber
of
cycles>
I
I
I-
I
...
6
1
...
90
Up to
4
ramp functions can be linked together. In order to do
this a value
>
90 must be entered for the cycle number of the
ramp function
1.
The following ramp functions must have the
value
1
for the cycle number. The number of ramp repetitions
can be selected between 1 and 9, according the cycle num-
bers between
91
and 99.
3.
A
segment number will automatically be assigned to
the programmed segments. Now you can enter the
following parameters for each
segment number:
the
start
temperature of the segment
the
end
temperature of the segment
the
run time
of the segment in seconds
(21)
Example:
segment no.
1
start
=
23.0°C
end
=
27.0°C
run time
=
600 sec
W-SP-1-23.0-27.0-600
All other segments
can be defined accordingly. Please note
that the temperature program is continual for all defined seg-
ments. This means that the end temperature of segment
i
must be the start temperature of segment i+l
.
4.
Start the program by entering the respective program no.
(This START command corresponds to the START of
the ramp via the menu function):
This will transfer the program from the
PC
to the circulator.
The
PC
can now be switched off or be used for other tasks.
The interface cable can be unplugged from the circulator.
After switching off the circulator the programs are lost.