Specifications

Serial Interface
-
1
NAMUR
1
Standard
I
Extended
I
I
I
I
/
Ramp parameters
1
With the
extended set of carnmands
a temperature program with six ramp seg-
ments can be defined. For every segment four parameters must be set:
Segment number End temperature of the segment
Start temperature of the segment Segment time
These parameters can be entered separately or in one line.
Standard
'
-
"
=
blank
-
Extended
/
NAMUR
/
Command
After selection of the segment number, the start and end temperature and the
segment time can be entered in any order.
Fault
F123
=
"RANGE
ERROR"
If
a value is entered that is out of the temperature limit, error message
"F123
appears. After that enter a correct value.
I
Select
segment nummer
I(no.:
1-6)
Set
segment time
(in
s,
resolution
0,01
s;
minimum
0
s,
maximum 86,400 s
=
24
hour)
Call up parameters of segment
<no.>
W-SS<va/ue>
W-SE<value>
Set
all parameters
of
segmentcno.>
-
1
Set
start temperature
-
I
Set
end temperature
Run
segment
Run to start temperature
(segment
no.:
1-6)
The start and end temperature of the segment (in OC, resolution 0,0I0C) must
be between the high and the low limit temperature of the circulator (see
13.6).
Before starting a segment first enter
W-ASxx.
Then the circulator runs to the start
temperature of the segment.