Specifications

Fault
Displays
I
ALARM
/
15.4
Sensor breakage
or
short circuit
This fault can correspond to the internal sensor
(RF
internal) or to an external sensor
(RF
external) con-
nected to the socket on the rear.
I
ALARM
/
-
lmprove shielding of the sensor connector cable.
(
see chapter
17.5
)
The sensor must be tested and possibly
exchanged
by
qualified service personnel.
4320°C
The message
"REF
means that the reference resistor
for the measuring value evaluation is out of order.
-
lmprove shielding of the sensor connector cable.
(
see chapter
17.5
)
return unit for servicing.
15.5
Undefined fault
I
ALARM
I
This can be caused by fault which only occurs for
a
short period of time, i.e. with a fluctuating bath level
when the filling level is very close to minimum.
Before returning the unit, top up with heat transfer liq-
uid. This fault can often be remedied in this
way!
In all other cases this unit must be checked by qualified
service personnel.
15.6
External
fault
RS232C
L-----A
Check the external system.
ALARM
-.
RS
2326
The circulator has been switched to fault status via the
interface.