Specifications

Operating
Operating
Switching on
Set the excess temperature limiter well above the
desired set temperature.
Switch on the unit via the mains switch
@
at the tem-
perature control unit
FIN.
Two temperatures are shown on
the main display:
set temperature (e.g. 43.20°C) in the top right corner
(the temperature momentarily to be reached)
actual temperature (e.g. 24.40°C) in the center.
FUM 43.20°C
1
This results
in
the following display if 'AUTOSTART: OFF'
a
START flashes in the display
MAlN MENU
0
I
wl
T-SELECT
CONFIG
CONTROL
[TI
T-SELECT
1
CONFIG CONTROL
(
MAlN MENU
1
I]
T-SELECT
CONFIG CONTROL
Select the main menu, alter the configuration, control
or temperature setting if desired and initiate START via
the ENTER key.
The main menu display then shows STOP. The unit can
be rendered inoperable at any time
by
executing STOP
via pressing the ENTER key.
!
C35:
In case of restarting the unit within approx.
5
min-
utes after it had been stopped the cooling com-
pressor will be switch on delayed for safety rea-
sons.
!
All refrigerated circulators:
To increase the life of the refrigerated unit,
it
is re-
commended to wait five minutes after switching off
before the unit is restarted.
This results in the following display if 'AUTOSTART:
ON'
The unit begins to work immediately. The pump, heat-
ing, cooling and control functions are active straight
away.
The unit can be rendered inoperable by selecting
STOP in the main menu and then pressing the ENTER
key.