3.4.8 FULL SCALE VALUE
This is the maximum capacity of the vessel IN THE ENGINEERING UNITS CHOSEN IN Section
3.4.3.
Press the CALIBRATION key to get to the CALIBRATION menu, then 4 for FS Value.
The unit prompts for the maximum capacity of the vessel in the already chosen engineering
units. Enter the vessel maximum value then press Enter.
3.4.9 ENABLE ALARM RELAY CONTROL
If the Alarm Relays are being used, now that calibration has been successfully completed,
the Alarm Relay control may be re-enabled.
Press the CONTROL key to get to the CONTROL menu, then 1 for Relay Settings then 5 to
Enable.
The right, bottom corner of the display should show ENA for enabled.
Press the DISPLAY key to go back to the Display menu.
THIS COMPLETES THE SETUP AND CALIBRATION PROCEDURE
NO FURTHER SETUP OR CONFIGURATION IS REQUIRED
THE NEXT SECTION (4) IS FOR REFERENCE ONLY