Select “Settings” from Main Menu. Select “Delete Videos”. A-360 will
prompt to conrm if deleting all les is desired.
19. VIEW SYSTEM INFORMATION
Select “System Info” from Settings Menu.
20. TROUBLESHOOTING
Make sure that A-360 battery is properly charged.
Make sure that the last Internal Software version is installed.
If Wi-Fi connection to iPad/iPhone is not working properly disable it
and work only with A-360 screen.
If the A-360 cannot be shut down try a hard reset by pressing the
Reset Button, located near the lens of the camera, using a thin bar.
Press and hold the On-Off button until “updating” is shown on the
A-360 screen (it will take around 20 seconds). Once the updating
process is nished (it can take seconds or a few minutes depending
on the update being performed) the screen will display “OK”
message.
Disconnect the cable between PC and A-360. Turn on A-360 and
check “System Info” to conrm the new software version is installed.
22. CLEANING INSTRUCTIONS
The A-360 is a non-critical device. Recommended cleaning is low-
level disinfection between each patient use. Be sure to follow your
institution’s specic cleaning procedures in consultation with this
manual.
1. Remove A-360 from the Airtraq and turn it OFF.
2. Cleaning: Use clean cotton gauze pads that are saturated with the
cleaning solution to wipe down the exterior surfaces of the A-360.
Use soft brushes with the cleaning solution to remove any residues
from areas that cannot be reached with the gauze pads. Be sure to
remove all visible contaminant or foreign matter. Be careful to keep
running liquid off the surfaces. The following cleaning solutions
may be used:
a. Enzymatic Cleaning Solutions (e.g.: Enzol
TM
Enzymatic
Detergent).
b. Neutral pH soap and water.
c. Sodium bicarbonate solution (8- 10 %).
3. The following disinfection agents may be used:
a. Peracetic Acid Solution (0.08%).
b. Isopropyl Alcohol (70%).
c. Solution containing 70% isopropyl alcohol and 2%
chlorhexidine (e.g. Clinell Wipe).
d. Solution containing chlorine dioxide (e.g. Tristel Wipe).
4. Blot dry the A-360 using an individual sterile surgical towel.
Caution:
• Do not autoclave.
• Do not rinse under running water.
21. UPDATING A-360 INTERNAL SOFTWARE
A-360 must be at least over 90% charged before upgrading.
Otherwise the unit will not upgrade.
Download latest version of A-360 Internal Software “A360SW_
Vxx.x” from downloads section in Airtraq website www.airtraq.com
and unzip it.
Turn On A-360 and connect to PC using the USB/Micro USB cable
provided (directly or through docking).
Open A-360 and go to folder “System”. Copy “A360SW_Vxx.x”
from PC to “System”.
Keep A-360 cable connected to PC and, using operating system
menus, remove safely A-360 from PC (as usually done to remove any
USB external device).
RESET