User's Manual
8 9
• A-360 will automatically turn on and start in Live Video mode.
• Slide lock system button into “Closed” position.
To prepare the A-360 for intubation.
• Flip and rotate the screen as desired depending on how the
intubation will be performed. Screen rotates 180
o
on one axis,
plus 270
o
on the perpendicular axis.
• Image can be rotated upside down when necessary using the
Rotate Icon
.
To remove A-360 from the Airtraq once intubation is nished:
• Place lock system button in “Open” position.
• Pull A-360 away from Airtraq.
7. TURN ON/OFF. STAND BY
A-360 can be started in 2 different ways:
• Automatically by sliding A-360 onto an Airtraq. It will log on to
“Live Video” function.
• Pressing for 2 seconds the orange On/Off button. A-360 will
log on to “Main Menu”.
A-360 starting process takes around 12 seconds, during which a
progress bar is shown on the screen.
A-360 will Turn Off by pressing for 2 seconds the orange On/Off
button. A-360 will prompt a screen that asks the user to conrm
shutdown.
A-360 will enter into Stand-By status only when A-360 is not
assembled onto the Airtraq, it is not in “Live Video”, and there is no
action from the user for 5 minutes. To reset from Stand-by, press the
On/Off button. A-360 will turn off automatically after 5 minutes in
Stand-by.
8. TOUCH PANEL & MENUS
Press the action icons on the screen to select an action (touch panel
also works when action icons are pressed using gloves).
Pressing the Back Icon will make A-360 to return to the previous
screen or menu.