User's Manual
4 5
Battery charger is supplied with different plugs. Install the appropriate
one.
Always use the battery charger provided with A-360.
3.3. Docking Station
Accessory to hold A-360 while not in use, A-360 can be locked to
the docking station. It protects the lens of the A-360 from dust.
The battery charger can be connected to the docking station.
4. A-360 SWITCHES, INDICATORS AND CONNECTORS
A. Mounting Lock System: To attach and remove A-360 from Airtraq
and from Docking Station. Slide button to place it in open or close
position.
B. Power Button: Push to turn On and Off.
A
B
C. Battery Charge and Power On Indicators: Turn green when on
charge. Left indicator turns red when the A-360 batteries are below
10% of its capacity.
D. Battery charge connector: Micro-USB Female connector to connect
to PC, battery charger or docking station.
5. A-360 BATTERY CHARGE
THE A-360 DOES NOT WORK WHEN IT IS BEING CHARGED.
IT SHOULD BE CHARGED AT LEAST 1.8 M AWAY FROM THE
PATIENT.
The A-360 is supplied with its battery discharged. It is recommended
to perform a full charge/discharge cycle of the battery before starting
to use the A-360 for clinical purposes.
After a full charge, the A-360 can operate, under standard
conditions, for around 100 minutes. Recharging takes approximately
150 minutes.
Without use, a fully charged battery, will hold 20 % of charge (20
minutes of use) after 20 days.
If the A-360 can be turned on then a minimum of 15 minutes of
operating time is guaranteed.
C
D