IV. 6. Technical characteristics
Dimensions: D x W x H: 368 x 419 x 699 mm.
Weight: 25.0 kg.
Noise level: ± 52 dBA
Flow values:
0 - 8 l/min.
Average oxygen content:
• at 2 l/min. 92 %.
• at 7 l/min. 91 %.
• at 8 l/min. 90%.
(values at 21 °C and at an atmospheric pressure of 1013 mbar). Max. recommended ow:
8 l/min. The maximum outlet pressure is 138 kPa. It is not recommended to use the NewLife
®
Intensity at ow values less than
2 l/min. unless utilized with the Pediatric Option Procedure.
Electrical power supply:
• 230 V - 50 Hz Europe / 115 V – 60 Hz / 230 V – 60 Hz (other countries,
depending on version)
Use the cable provided, and check that the electrical characteristics of the mains socket used
match those indicated on the manufacturer’s plate on the rear panel of the machine.
• Mean power rating: 410 watts
Class II
Type B
2.0 amps (230 V).
4.0 amps (120 V).
Filters:
At the rear of the device: a dust lter.
Before the oxygen outlet: a product lter < 2.0 µm.
Air circulation:
A fan cools the compressor compartment.
Environmental limit conditions:
The performances of the device (especially the oxygen concentration) are quoted at 21 °C and
1013 mbar. They may change with temperature and altitude.
- The device should preferably be stored and transported in the vertical position.
- The device must only be used in the vertical position.
- Ambient temperature of between 5°C and 40 °C (operation).
- Storage temperature range from -20 °C to 60 °C.
- Relative humidity of between 0% and 95% (operation and storage).
- IPX1: Protected against dripping water (complies with the EN 60601-1 standard)
7-GB