II. STARTING UP / INSTALLATION
II. 1. Use in direct oxygen therapy.
a - Ensure that the switch (1) is in the 0 position.
b - If used with a humidier:
Unscrew the ask and ll it with water up to the line (see the humidier
instructions). Then screw the humidier ask onto its lid until there are
no leaks from it.
c - Connect the oxygen administration tube to the humidier outlet
nozzle or connect the administration nasal cannulas onto the
concentrator. The tube between the patient and the NewLife
®
Intensity should be less than 60 metres long, in order
to ensure that the oxygen ow rate remains satisfactory.
d - Ensure that all of the parts are connected correctly so as to avoid
leaks.
e - Plug the power cable into a mains socket.
f - Press the switch to the start position - I . An alarm operating test
is carried out automatically when the machine is switched on (this
test lasts about 5 seconds).
g - Turn the ow adjustment knob (6) to the prescribed value.
The prescribed valve line should appear to split the middle of the
owmeter ball.
h - Check that the oxygen ows out of the administration device
(nasal cannulas or other) by placing the orice(s) on the surface of a
glass of water. The ow should disturb the surface of the water.
i - Adjust the nasal cannulas or mask to suit your face.
Remark: the optimal oxygen concentration is obtained about ten minutes after the device is
switched on (90 % of the concentration is obtained after around 5 minutes).
At the end of the treatment, press the push button (1) to place it in the 0 position to stop the device.
The oxygen enriched air ow continues for approximately 1 minute after the device is stopped.