Owner's manual

28
MN138-1 Rev H
02/14
5.3 Troubleshooting Chart
Problem
Probable Cause
Solution
Unit does not run.
I/0 power switch in “I” position.
• No power to unit.
• Unit circuit breaker tripped/faulty
• Faulty electrical connections.
• Faulty circuit board.
• Faulty I/0 power switch.
• Check wall outlet for power.
• Reset or replace circuit breaker.
• Check electrical connections.
• Replace circuit board.
• Replace I/0 power switch.
Compressor runs with intermittent
alarm. Oxygen concentration at max
flow rate is within specifications.
• Faulty pressure transducer.
• Replace circuit board.
Compressor runs and shuts down
periodically. Only start up alarm
activates
• Faulty oxygen monitor circuit board.
• Replace oxygen monitor circuit board.
Compressor runs and shuts down
periodically with 5-beep intermittent
high temperature alarm.
• Restricted air flow.
• Unit overheating due to improper
location. Compressor thermally cut off
due to excessive heat. NOTE: It will
not restart until it cools down.
• Faulty cabinet fan.
• Temperature probe wire harness faulty
or disconnected from circuit board.
• Remove obstruction.
• Locate unit away from heating source,
providing adequate ventilation on all
sides.
• Do not operate more than one unit
side by side.
• Replace cabinet fan.
• Replace or reconnect temperature
probe.
Compressor runs with 3-beep
intermittent low pressure alarm
and low oxygen concentration.
!
• Leak !
• Reduced air intake
• Weak compressor
• Faulty circuit board.
• Leak test and repair leak.
• Check compressor intake path for
obstruction. Clean/remove
obstruction.
• Replace compressor.
• Replace circuit board.
Compressor does not start. I/0
power switch in “I” position,
intermittent 3-beep low pressure
alarm, and cabinet fan turns.
• Extreme cold start.
• Faulty electrical connection for
compressor.
• Faulty capacitor.
• Faulty circuit board.
• Allow unit to reach room
temperature.
• Check electrical connections for
compressor.
• Replace capacitor.
• Replace circuit board
4-beep high pressure alarm is
activated and both compressor
and valves do not operate.
• Restriction in exhaust muffler.
• Contaminated sieve beds.
• Faulty solenoid valve.
• Faulty circuit board.
• Replace exhaust muffler.
• Replace sieve beds.
• Replace solenoid valve manifold.
• Replace circuit board.