Manual
AirSep
®
Corporation
NewLife
®
Elite Service Manual
NOTE
If a pressure transducer circuit board,
cut tie-wrap at circuit board and
disconnect green 1/4-inch tube. Plug/
cap open end of tube until new circuit
board is replaced.
4 Push in on the locking tab to the alarm 6-pin connector to
disconnect it.
5 With a slotted screwdriver, push in on the board support tabs,
while you lift each corner of the circuit board.
6 Remove the circuit board.
NOTE
Handle the new circuit board only
by the edges to prevent electro static
damage to the unit.
5.7.2 Circuit Board Installation
NOTE
If a pressure transducer circuit board,
rst reinstall 1/4-inch green tube with
new tie-wrap.
1 Push the circuit board on to the support tabs.
2 Firmly plug in the 6-pin connector to the circuit board so that
the connector’s locking tab locks against the circuit board.
3 Plug in the 10-pin connector.
4 Reconnect the back and side panels.
5-9