4 Install the new lter with the inlet side in the same position as
before. Push the tubing together so that the tubing overlaps
the product lter connections, and secure it with the tie-wraps.
5 Record information about the product lter replacement on
the History Record Label, which is discussed in Section 4.1.5.
6 Reconnect the back and side panels.
4.1.3 Optional Filter Replacement
The optional internal felt lter requires changing every 5,000 hours of
use. See below on instructions for changing the lter.
1 Set the unit’s ON/OFF switch to the OFF position, and unplug
the power cord.
2 Remove right side panel to locate the felt intake lter.
3 Remove lter in the unit, and replace with a new lter.
4 Remove left side panel and record information about the lter
replacement on the History Record Label.
5 Reconnect both side panels.
4.1.4 Battery Replacement
Each time the NewLife unit is turned on, the alarm must sound loudly
for approximately ve seconds to indicate a good battery. An alarm
that does anything other than sound loudly for ve seconds indicates
a weak battery and requires replacement.
To replace the battery, take the following steps:
1 Set the unit’s ON/OFF switch to the OFF position, and unplug
the power cord.
2 Remove the left side panel.
3 Lift the battery out of the battery holder.
4-2