Instruction Manual

22
Cleaning, Care, and Proper Maintenance
Compressor
The compressor must be replaced after approximately 3,000 hours of use. When you remove the air intake/gross
particle filter for cleaning, take note of the reading on the hour meter positioned beneath this filter. (See Figure 8.) When
the reading approaches 3,000 hours, contact your Equipment Provider for replacement of the compressor. At this
time, the Equipment Provider should also replace the product filter and felt filter.
Figure 8: Hour Meter Position
Hour Meter