Owner's manual

GB
NewLife
®
Elite
III.
CLEANING
-
MAINTENANCE
III.
1.
Cleaning
Only the outside of the
NEWLIFE
ELITE
is to be cleaned, with a dry cloth or,
if
necessary, a damp sponge and clean or soapy water, then thoroughly dried with
wipes
and an alcohol based solution. Acetone, solvents or any other flammable
products
must
not
be
use
d. Do not use abrasive powders.
The removable dust filter (9) must be cleaned in
soapy water after approximately 100 hours of use, or
when
required. Rinse and wipe dry. Fit a dry filter.
III.
2.
Everyday
disinfection
Due to the presence of the product filter inside the
device, everyday disinfection only concerns the external oxygen therapy acces-
sories: humidifier, probes, nasal cannulas (refer to the respective instructions
for use).
The
use
of
alcohol
based
solutions
means
that
the
device
must
be
switched
off.
a
-
The
following
minimum
guidelines
must
be
respecte
d:
Humidifier :
Daily:
- empty the water from the humidifier.
- rinse the humidifier flask under running water.
- Fill the humidifier up to the mark with slightly mineralised water.
Regularly:
- Disinfect the equipment by immersing them in a disinfectant solution (in
general, we recommend using water containing a small amount of chlorine
bleach).
- Rinse and dry.
- Check that the humidifier lid seal is in good condition.
Oxygen administration devices:
Follow the manufacturers instructions.
b
-
For each
new patient:
The humidifier must be sterilised if possible or changed. The
NEWLIFE
ELITE
must be cleaned and disinfected as per the above instructions. The dust
filter
and the product filter should preferably be changed. The entire oxygen adminis-
tration circuit (oxygen therapy nasal cannulas, etc.) must be changed.