Product Manual

MN078-1 Rev K
41
[Read the Important Safety Rules section before operating this equipment.]
Setting the Pediatric Flow meter
When using the pediatric flow meter, the unit will not reach proper concentration at the pediatric setting (less than 2 lpm) until
you bleed off a portion of the oxygen by opening the primary flow meter (on the left side of the unit). Follow the procedure
below when using the pediatric flow meter.
1. Follow the start-up instructions 1-7 as outlined on page 9 & 10.
2. Set the pediatric flow meter to the prescribed flow.
3. Set the primary flow meter to 2 lpm to bleed off excess product, and allow the unit to achieve maximum concentration.
The NewLife Elite is appropriate for usage by two patients,
provided the combined flow is a minimum of 2 lpm and does
not exceed the maximum capacity of the concentrator.
The NewLife Elite Oxygen Concentrator must be operated
for at least five minutes at 2 lpm before using the unit.