Product Manual
MN078-1 Rev K
29
EMC SUMMARY: PART III FCC EMC EMISSION TEST RESULTS
FCC PART 15,
SUBPART B
TEST REQUIREMENTS
MARGIN BELOW (-) / ABOVE (+) THE
LIMITS
COMPLIANCE
(YES / NO)
Source Testing: ULTRATECH GROUP OF LABS File #: AIR095-FCC15B (Elite 120V 60Hz) 9Aug2010
15.107(a) ,
CLASS B
AC Power Line Conducted Emissions
Measurements
-28.3 dB @ 13.007 MHz
YES
15.109(a) ,
CLASS B
Radiated Emissions from Unintentional
Radiators (Digital Devices)
-1.5 dB @ 80.00MHz
YES
Source Testing: ULTRATECH GROUP OF LABS File #: AIR-116_FCC15B_Elite 230v 50Hz
15.107(a) ,
CLASS B
AC Power Line Conducted Emissions
Measurements
- 9.3 dB @ 0.156 MHz
YES
15.109(a) ,
CLASS B
Radiated Emissions from Unintentional
Radiators (Digital Devices)
- 7.1 dB @ 112.00 MHz
YES