Product Manual
MN078-1 Rev K
27
EMC SUMMARY: PART I IMMUNITY TEST RESULTS Continued
Standard
Description
Severity Required
Performance
Criteria Met
Performance
Criteria Allowed(3)
IEC 61000-4-11
EN 61000-4-11
For 230V 50 Hz
Dips
Dips
Dips
Interrupts
< 5% UT (>95% dip in UT) for 0.5 cycles
40% UT (60% dip in UT) for 5 cycles
70% UT (30% dip in UT) for 25 cycles
< 5% UT (>95% dip in UT) for 5 seconds
* UT: a.c. mains voltage prior to
application of the test level.
A
A
A
C
A
A
A
C
IEC 61000-4-11
EN 61000-4-11
For 120V 60 Hz
Dips
Dips
Dips
Interrupts
30% - 25 cycles for 3 times
60% - 5 cycles or 3 times
>95 - 0.5 cycles for 3 times
>95 - 250 cycles for 3 times
A
A
A
C
A
A
A
C
IEC 61000-3-2
EN61000-3-2
Harmonic Current Emissions
Class A
N/A
N/A
IEC 61000-3-3
EN61000-3-3
Voltage
Fluctuation and
Flicker in Low-
Voltage Supplies
Voltage Fluctuation
Flicker
N/A
N/A
N/A
N/A
EN 60601-1-2
CLAUSE
36.202.14
(230V 50 Hz)
(120V 60 Hz)
Variations of
power frequencies
Variations of power frequencies
at +/- 1Hz (49 Hz and 51Hz)
at +/- 1Hz (59 Hz and 61Hz)
PASS
PASS
PASS
PASS