Product Manual

MN078-1 Rev K
23
Product Specifications Continued
Excessive Temperatures**
ISO 8359:7.1
IEC 60601-1: 11.1
Maximum O
2
Outlet Temperature Rise
+ 5.4
o
F (3
o
C)
Flow Rate Range
ISO 8359 Section 8_50.3
IEC 60601-1: 7.4
1% maximum difference in flow rate of allowable +/- 10%
O
2
Concentration
at extreme voltage conditions
ISO 8359 Section 8_50.4
IEC 60601-1: 7.4
Voltage extreme % Volume
132V 94.6
103V 94.0
253V 91.0
195V 90.0
ISO 8359 Section 8_50.5
Mean O2 Concentration
IEC 60601-1: 7.4
Voltage extreme % Mean O
2
% Individual Deviation from mean
(+/- 3% allowable)
132V 94.6 0.6
103V 94.0 0.6
253V 91.0 0.6
195V 90.0 1.4
ISO 8359 Section 8_50.6
Mean Flow Rate
IEC 60601-1: 7.4
Voltage extreme % Mean Flow % Individual Deviation from mean
(+/- 3% allowable)
132V 4.87 0.8
103V 5.11 0.8
253V 4.99 1.4
195V 5.00 1.4