The Manufacturer does not recommend the sterilization of this equipment.
If the unit has not been used for an extended time period, it needs to operate for a minimum of minutes before
power failure alarm can become activated.
For units equipped with the oxygen monitor option - When you turn the unit on, it’s normal for the amber
OXYGEN MONITOR light to turn on and remain on for up to five minutes.
For prescriptions levels below 1 lpm, you must operate the oxygen concentrator for at least 5 minutes at 2 lpm.
The concentrator releases warm air out the bottom of the unit which can permanently discolor temperature
sensitive flooring surfaces such as vinyl. The concentrator should not be used over flooring that is sensitive to
heat staining. The Manufacturer is not responsible for flooring that becomes discolored.
To prevent a void warranty, follow all manufacturers’ instructions.
Do not attempt any maintenance other than the possible solutions listed within the manual.
Portable and mobile RF communications equipment can effect medical electrical equipment.
There is never a danger of depleting the oxygen in a room when you use your Oxygen Concentrator unit.
The standard NewLife Elite Oxygen Concentrator accommodates prescriptions from
1 lpm minimum to 5 lpm
maximum.
AirSep offers the OxySafe as an optional accessory. This is intended to be used in conjunction with the NewLife Elite
concentrator. For customers in regions requiring compliance to EN ISO 8359:1996-Ammendment1:2012, this accessory
will meet this need.
The OxySafe is a thermal fuse to stop the flow of gas in the event that the downstream cannula or oxygen tubing is ignited
and burns to the OxySafe. It is placed in-line with the nasal cannula or oxygen tubing between the patient and the oxygen
outlet of the NewLife Elite
.
For proper use of the OxySafe, always refer to the manufacturer’s instructions (included with each OxySafe kit).
AirSep offers an OxySafe kit that includes an OxySafe, 2” of tubing, and a DISS connector to connect to the
oxygen outlet.; PN 20628668