MAINTENANCE
Maintenance Procedures
The ScanX is designed for many years of trouble-free operation. Maintenance as
described herein is minimal.
IMPORTANT: Do not spray solvents or liquid directly on the scanner.
Cleaning the ScanX
Turn off the ScanX disconnect the line cord from the Mains wall outlet and dis-
connect the computer connection cable from the ScanX before cleaning. Wipe
the outside surfaces with a soft paper towel dampened with a disinfectant solu-
tion or non-abrasive household cleaner. Be careful not to allow solvents TO RUN
OR DRIP into the ScanX. This could cause damage to the ScanX. Allow to air dry
before plugging in or turning back on.
Note: ScanX Cleaning Sheets cannot be used on the ScanX
®
Intraoral
(P/N B7100) because of height limitations
.
Cleaning the Plate Transport
Over time, small debris and dust may accumulate in the plate transport mecha-
nism causing a loss in image quality and possible damage to the PSPs.
To ensure optimal performance of the ScanX
®
(P/N B7200) and the Ortho
ScanX
®
(P/N B7250), the plate transport should be cleaned once per week using
a new ScanX Cleaning Sheet each time. Sample sheets are included with the
ScanX and additional sheets can be purchased from your dealer. Refer to
ACCESSORIES page 27.
Disinfecting the ScanX
No disinfection is required for the ScanX.
Cleaning the Plate Guides
Wash the Plate Guides with a non-abrasive cleaner and water. Dry thoroughly.
IMPORTANT: Plate Guides must not be put in an autoclave.
Note: Always use standard infection control procedures when handling
devices that contact the patient.
Disinfecting the Plate Guides
There is no reason to routinely disinfect the Plate Guide unless contamination is sus-
pected. If a Plate Guide has touched a contaminated surface, it may be immersed
briefly in a cold sterilant according to sterilant manufacturers directions. After dis-
infection, clean and dry the Plate Guide using the instructions above.
POWERING DOWN THE SYSTEM
The ScanX is designed to be left on continuously during the active day. At the
end of the day, or whenever desired, power down the system simply by pressing
and holding the membrane READY switch ( ) on the Keypad Control Panel
for approximately two seconds, until the green LED above the READY switch
extinguishes.
IMPORTANT: Never power down the system during a
scanning session.