User`s manual

TTRROOUUBBLLEE SSHHOOOOTTIINNGG
PPRROOBBLLEEMM PPOOSSSSIIBBLLEE CCAAUUSSEE PPOOSSSSIIBBLLEE SSOOLLUUTTIIOONNSS
1. Film too light. a. Developer exhausted. a. Change chemistry after 300-
350 films or 2 weeks, whichever
comes first.
b. Chemistry does not reach b. Call your authorized Air
factory set temperature. Techniques dealer for service.
c. X-ray machine is c. Call your x-ray dealer for
improperly set or inconsistent. service.
d. Developer is contaminated. d. Dispose of chemistry and clean
with Formula 2000. Replace
chemistry and water.
e. Improper or outdated film. e. Check date on film box.
f. Not using specially f. Use Peri-Pro chemistry to
formulated Peri-Pro assure film quality and archival
chemistry. quality.
2. Film too dark or a. Developer temperature is a. Call you authorized Air
grainy. too high. Techniques dealer for service.
b. X-ray machine is b. Call your x-ray dealer for
improperly set or inconsistent. service.
c. Film has been exposed to c. Check for and correct light
light. leaks in the dark room or
daylight loader. Place the inlet
cover on the film inlet before
leaving the darkroom or before
removing hands from the
daylight loader cuffs.
d. Not using specially d. Use Peri-Pro chemistry to
formulated Peri-Pro assure film quality and archival
chemistry. quality.
3. Film is cloudy or a. Fixer is exhausted. a. Change chemistry after 300-
has brownish surface. 350 films or 2 weeks, whichever
comes first.
b. Fixer temperature is too b. Wait until the ready light
low. stays on continuously, indicating
that the set-point has been
reached.
c. Chemistry does not reach c. Call your authorized Air
factory set temperature. Techniques dealer for service.
d. Improper or outdated film. d. Check date on film box.
Make sure that the film being
used is recommended for
automatic processing.
11
NOTE: Dispose of any chemicals in accordance with local codes.NOTE: Dispose of any chemicals in accordance with local codes.