Brochure

This Product Exceeds OSHA, ANSI and Other International Standards.
p:5
Ductless Fume Hoods 24 • 36 • 48Product Filtration
MULTIPLEX FILTRATION SYSTEM, SUMMARY
Pre-Filter Main Filter
Electrostatic Protects the main filters from aerosols, mists, dust and particulates with filter efficiency
superior to 95.5% down to 0.5 microns.
Standard --
Activated Carbon FILTCO
Sourced. A single carbon filter containing activated carbon granules chemically
formulated to capture one or more specific vapors or family of vapors.
Single:
One type of activated carbon.
-- Specify
Stacked:
Two or more single filters each
with a different type of carbon.
-- Specify
HEPA/ULPA A self-contained filter designed to physically capture particles larger than 0.3 microns (HEPA)
or 0.12 microns (ULPA). When used with a HEPA/ULPA filter the ductless fume hood may be
applied as a Class I Biological Safety Cabinet.
-- Specify
Purair P5-36, shown with
Multiplex Filtration System.
The Purair Basic Series
ductless fume hood
maintains a constant
face velocity of 100 FPM in
compliance with USA and
international standards
for safety and performance.
Contaminated air is pulled
through the Multiplex
filtration system where
activated carbon adsorbs
chemical vapors and/or
particulates if HEPA/ULPA
filters are used. Clean
air is returned
to the room.
A. The main filter is
easy to replace, no tools
required. The filter clamps
tightly against the filter
gasket to prevent filter
bypass and maintain
filter integrity.
Room Air Intake
PURAIR
®
AIRFLOW PATTERN
Fan / Blower
Exhaust Air
Filtration
A
Main Filters
of carbon and/or
HEPA/ULPA
Electrostatic Pre-filter
Pre-filter may be
changed while unit
is running
The Multiplex filter
configuration permits a
customized combination
of filter media for a broad
range of chemical fami-
lies and biological agents
if required. EFT Filtration
Technology broadens the
Air Science application for
ductless fume hoods.
PURAIR BASIC
FEATURES & BENEFITS
Purair Basic products
are available in 3
standard sizes, each
with 4 configuration
options and metal or
polypropylene con-
struction, totaling 24
standard models.
High capacity air
handling system
delivers face velocity
of 100 FPM.
A low airow alarm
warns of insufficient
face velocity.
The Air Science filter
assembly is easy
to access, easy to
change.
A unique filter
clamping design
eliminates bypass
leakage outside
the cabinet.
Accessories include
an optional filter
saturation alarm.