User Manual
  2 
Specification 
Input Specifications 
Input A 
Configurable options 
  Input coupling: 
  Input impedance: 
  Attenuation: 
  Active edge: 
  Low pass filter: 
  Trigger threshold: 
AC or DC 
1MΩ or 50Ω 
1:1 or 5:1 
Rising or falling, or width high or low 
Filter In (~50kHz cut-off) or Out 
Variable threshold for both DC and AC coupling 
Input Impedance:  
1MΩ//25pF (DC or AC coupled) 
or 50Ω nominal (AC coupled only). 
Frequency Range: 
< 0.001Hz to >125MHz (1MΩ, DC coupled). 
< 30Hz to >125MHz (1MΩ, AC coupled). 
< 500kHz to > 125MHz (50Ω, AC coupled). 
Trigger Threshold: 
  DC coupled: 
  AC coupled: 
0 to 2V (1:1 attenuation) or 0 to 10V (5:1 attenuation). 
Average ± 50mV (1:1 attenuation) or ± 250mV (5:1 attenuation). 
Sensitivity:  Sinewave - 15mVrms 30Hz to 100MHz, 25mV to 125MHz 
at optimum threshold adjustment. 
Input B 
Input Impedance: 
50Ω nominal (AC coupled). 
Frequency Range:  < 80MHz to >3GHz. 
Sensitivity:  12mVrms 80MHz to 2GHz, 25mVrms to 2.5GHz, 50mVrms to 3GHz. 
Maximum Input Signal:  < 0dBm recommended, +13dBm (1V
rms) maximum. 
Input C 
Input Impedance: 
50
Ω
 nominal (AC coupled) in-band. 250k
Ω
 at DC. 
Frequency Range:  < 2GHz to >6GHz (typically 1.8GHz to 7.5GHz). 
Sensitivity:  25mVrms (–19dBm) 2GHz to 6GHz. 
Maximum Input Signal:  < +16dBm (1.5Vrms) recommended. Damage level +25dBm (4Vrms). 
External Reference Input 
Input Impedance: 
>100k
Ω, 
AC coupled. 
Frequency:  10MHz. 
Signal Level: 
TTL, 3V
pp
 to 5V
pp
 CMOS or 1 to 2Vrms sinewave. 
Maximum Input Voltage 
Inputs A, B, C, and 
External Reference: 
30V
DC; 30Vrms 50/60Hz with respect to earth ground   
Note that the inputs will not be damaged if subjected to an accidental short-term connection to a 50/60Hz 
line voltage not exceeding 250V
rms, or 250V DC. 