Instructions
Page 73 
Zero Span Audio Demodulation 
Demod. Modes: 
AM or FM 
Internal Audio: 
Internal loudspeaker with adjustable volume and mute 
Audio Out: 
30 mW into 32 Ohms mono or stereo headphones, adjustable volume, 
3.5mm jack socket (auto disconnect of internal loudspeaker) 
Audio Filter 
Selectable low-pass filter to attenuate high frequency interference. 
2 pole filter with turnover point at approximately 3 kHz. 
Zero Span Waveform Demodulation (only with Option U02) 
Display Types: 
AM Waveform, FM Waveform, Carrier Level or Carrier Waveform. 
Modulation Range: 
+/-5% to +/-100% full scale (AM) or +/-1kHz to +/-1MHz full scale (FM).  
Modulation Rate: 
35Hz to 100KHz. 
Waveform Trigger 
Variable trigger point on rising or falling edge. 
13.1.2  Amplitude Measurement (Level) 
Amplitude Range 
Display Range: 
84 dB typical 
Measurement Units 
dBm or dBµV (dB milliwatts or dB micro volts) 
Reference Level: 
Selectable between -40dBm (67 dBµV) and +20dBm (127 dBµV) with a 
resolution of 1dB 
Amplitude Accuracy 
Ref. Level Accuracy: 
Better than ± 1 dB at 2000 MHz at 10 dB below reference level 
Level Flatness: 
Better than ± 1 dB relative to 2000 MHz over 10 MHz to 6000 MHz  
[10 MHz to 3600 MHz] 
Amplitude Linearity: 
Better than ± 1 dB over 60dB range down from reference level 
Displayed Average Noise Level (DANL) 
DANL: 
Better than – 120 dBm (-13 dBuV) average displayed noise floor  
(ref. level = – 40 dBm (+67 dBuV), RBW = 10 kHz, VBW = 1 kHz, span 
1MHz) over 10 MHz to 5500 MHz [10 MHz to 3600 MHz] 
Noise per Hz: 
Better than -160 dBm per Hz equivalent 
Distortion and Spurii 
3rd Order Intermod.: 
< – 60dBc (typically -65dBc) for two signals at 10dB below the 
reference level, 500MHz and 502MHz. 
Harmonic: 
< –65dBc at 10dB below reference level (100MHz)  
Signal Images: 
<– 55dBc, typically <– 60dBc (Normal sweep, 300Hz to 3MHz RBW) 
Other Signal Related 
Spurii 
< – 60dBc for signals 10dB below the reference level. 
Residual Spurii: 
< – 70 dB below the reference level. 
Amplitude Markers 
Number of Markers: 
One, Two, or None 
Marker Resolution: 
0.1 dB 
Amplitude Readout: 
Display of absolute and difference amplitudes for both markers in 
graticule units (dBm or dBuV) or linear units (mV or uW). 
Amplitude Limits (only with Option U02) 
Limit Types: 
Limit Lines from numeric values, or Limit Patterns from Files. Pattern 
files are created using PSA-Manager software. 
Number of Limits: 
One or two limits on screen. Up to 999 patterns can be stored 
Limits Comparator: 
Conditions of Above, Below, Inside or Outside creating actions of 
beep, stop sweep, log sweep and pulse out  