Instructions
  - 18 - 
Marker>Centre moves the active marker to the sweep centre frequency. 
Move Left/Right moves the active marker left or right either in frequency 
steps or from peak to peak of the trace depending on the Marker Mode 
(Scroll or Peak Find). 
Note that the navigator hard keys (Left/Right) can perform a similar function. 
4.7   Setup/Functions 
4.7.1 Setup/Functions > Logging 
Logging functions are only available when Option U02 is installed. 
Log Type creates a pop-up menu that selects the type of data to be logged. Options 
are Centre Level, Peak Level, Full Trace, or Screen Image. 
Log Trigger creates a pop-up menu that selects the trigger for logging entries. 
Options are Timer, Manual (using Run/Stop key), External trigger input (+ve or -ve 
level change), Limits Event or Continuous (log after every sweep). 
Timer Setup sets the internal timer interval between 1 second and 99 minutes. Note 
that the actual logging interval will depend upon the sweep time which may be much 
longer. 
Logging Control opens a control menu that puts the instrument into logging mode. 
No changes to the instrument set-up can be made once logging is commenced. 
All logging conditions must be set prior to using the logging control screen.
The Logging control screen provides an overview of the logging set-up and has keys as follows: 
Change Default the auto-incrementing default name for log files can be changed to any five 
character name. 
File Utilities opens a file list screen for log files enabling renaming and deletion. 
Change Name enables a specific name for the new log file (rather than default name) to be 
entered. 
Enable Logging opens the new log file and enables logging entries to the file (a warning screen 
is displayed requiring confirmation). 
Pause/Resume logging entries can be paused, but no changes to the set-up can be made. 
Close File disables logging and closes the file. Once closed, no further entries can be made. 
Cancel/Exit exits the control screen if no log file is open. If a log file is open, it pauses logging 
and provides options to resume or to close and exit. 
4.7.2 Setup/Functions > Auto Measure 
 Automatic measurements requires option U02. Measurement of Channel Power 
(CP), Adjacent Channel Power Ratio (ACPR), and Occupied Bandwidth (OBW) are 
provided. 
Centre frequency, span and RBW must be set manually prior to measurement. 
On/Off toggles the automatic measurement on or off. When on, normal markers are 
suppressed and marker lines are positioned automatically. Measurement results 
are shown in the marker readout area. 
Mode creates a pop-up menu that controls the sweep mode (Single or Continuous) 
and the display mode (Results or Setup). 
Measure Type creates a pop-up menu from which the measurement type is 
selected. 
Set Channel Bandwidth appears for CP and ACPR only. Creates a numeric entry 
screen from which the relevant channel bandwidth can be entered. The main 
channel is shown between yellow lines. Channel power is displayed in dBm. 
Marker 
>Centre 
Move 
Left 
Move 
Right 
Logging 
Log 
Type 
Log 
Trigger 
Timer 
Setup 
Logging 
Control 
Auto 
Measure 
On/Off 
Mode 
Measure 
Type 
Set Chn 
B/W 