Instructions
Page 75
File Storage:
Scan files can be saved under automatic or user defined names and
recalled to the screen or exported to external programs.
13.1.5 Memory Storage and Data Logging
Memory Storage
Memory Type:
Internal non volatile Flash memory (1.8GB nominal).
Store Trace:
Up to 999 traces can be stored under either default file names or user
entered file names. Traces are stored as tables of amplitude versus
frequency and can be imported into other programs, as well as being
recalled to the screen. Traces can be stored from either the live trace
or the view trace and include State information.
Recall Trace:
Recalls any stored trace to the reference trace of the display. The
State can be recalled independently from the Trace
Store Screen:
This function copies the whole screen area (excluding key area) to
memory as a bit-map image. Up to 999 screens can be stored under
either default file names or user entered file names..
Recall Screen:
Recalls any stored screen image to the display.
Store Set-up:
Up to 999 instrument set-ups can be stored under either default file
names or user entered file names. All settings of the instrument are
saved.
Recall Set-up:
Recalls any stored set-up, overwriting the existing settings of the
instrument.
Store Scan:
Only with Option U02. Up to 999 scans can be stored under either
default file names or user entered file names. Scans are stored as
tables of amplitude versus frequency and can be imported into other
programs, as well as being recalled to the screen. Scans can include
State information.
Recall Scan:
Recalls any stored Scan to the display. The State can be recalled
independently from the Scan.
Store Logging File:
Only with Option U02. Up to 999 logging files can be stored for
download and analysis by PSA-Manager software.
Data Logging (only with Option U02)
Log Files
Files of up to 25,000 data entries can be created. Up to 999 files can
be stored. Files are read using PSA-Manager Software.
Logged Data:
Log files can be created from Peak Value, Centre Frequency Level,
Complete Sweep, or Screen Image
Log Trigger:
Log entry can be triggered by Key Press, Continuous (every sweep),
External Input, Internal Timer, or Limits Comparator
Timer Interval
Adjustable between 2 seconds and 100 minutes per entry
Time Stamp
Each entry is time stamped from the real-time clock
13.1.6 Connectors and Display
Signal Input
Input Connector:
N connector, 50 Ohms
VSWR:
1.5:1 typical
Maximum
Input Level:
Absolute Maximum Input Level:
+ 25 dBm or +132 dBµV (4V rms) or 50V dc
Other Connectors
Power Connector:
1.3mm dc power socket for AC power adaptor/charger as supplied.
USB link to PC
interface:
Mini USB connector (5 pin mini B) which acts as a USB “device” for
connection to a personal computer.