Instructions
- 22 -
In Scroll mode it moves in display point steps. In Peak Find mode it jumps between the peaks
of the zoomed display.
Pan to Marker re-centres the zoomed display to the marker position. The marker can be
scrolled off the graticule area but remains visible within the lower display.
The marker readout shows the centre of the group of frequencies represented by the display
point. No level information is shown until maximum zoom is reached.
Store/Recall scan files are saved into the TRACES folder and are identified by having a
starting character of $. A file can be stored under an auto-incrementing default name of
$SCAN001 or a user defined name of $ followed by up to 7 characters.
The file contains both the Data of the scan (frequency and level) and the set-up of the
instrument which is referred to as the State.
For file Recall, the user can choose to recall the Data & State together or either one alone. If
only the data is recalled, the upper annotation area may not be relevant to the recalled scan.
Build for Export to export a file for external use it is necessary to “build” the CSV file so that it
contains the scan data. The user can choose to create a CSV which has a header containing
the set-up information or to omit it so that it contains only the data.
The Build for Export function can be slow when the number of samples in the file is large.