Instructions

- 18 -
Marker>Centre moves the active marker to the sweep centre frequency.
Move Left/Right moves the active marker left or right either in frequency
steps or from peak to peak of the trace depending on the Marker Mode
(Scroll or Peak Find).
Note that the navigator hard keys (Left/Right) can perform a similar function.
4.7 Setup/Functions
4.7.1 Setup/Functions > Logging
Logging functions are only available when Option U02 is installed.
Log Type creates a pop-up menu that selects the type of data to be logged. Options
are Centre Level, Peak Level, Full Trace, or Screen Image.
Log Trigger creates a pop-up menu that selects the trigger for logging entries.
Options are Timer, Manual (using Run/Stop key), External trigger input (+ve or -ve
level change), Limits Event or Continuous (log after every sweep).
Timer Setup sets the internal timer interval between 1 second and 99 minutes. Note
that the actual logging interval will depend upon the sweep time which may be much
longer.
Logging Control opens a control menu that puts the instrument into logging mode.
No changes to the instrument set-up can be made once logging is commenced.
All logging conditions must be set prior to using the logging control screen.
The Logging control screen provides an overview of the logging set-up and has keys as follows:
Change Default the auto-incrementing default name for log files can be changed to any five
character name.
File Utilities opens a file list screen for log files enabling renaming and deletion.
Change Name enables a specific name for the new log file (rather than default name) to be
entered.
Enable Logging opens the new log file and enables logging entries to the file (a warning screen
is displayed requiring confirmation).
Pause/Resume logging entries can be paused, but no changes to the set-up can be made.
Close File disables logging and closes the file. Once closed, no further entries can be made.
Cancel/Exit exits the control screen if no log file is open. If a log file is open, it pauses logging
and provides options to resume or to close and exit.
4.7.2 Setup/Functions > Auto Measure
Automatic measurements requires option U02. Measurement of Channel Power
(CP), Adjacent Channel Power Ratio (ACPR), and Occupied Bandwidth (OBW) are
provided.
Centre frequency, span and RBW must be set manually prior to measurement.
On/Off toggles the automatic measurement on or off. When on, normal markers are
suppressed and marker lines are positioned automatically. Measurement results
are shown in the marker readout area.
Mode creates a pop-up menu that controls the sweep mode (Single or Continuous)
and the display mode (Results or Setup).
Measure Type creates a pop-up menu from which the measurement type is
selected.
Set Channel Bandwidth appears for CP and ACPR only. Creates a numeric entry
screen from which the relevant channel bandwidth can be entered. The main
channel is shown between yellow lines. Channel power is displayed in dBm.
Marker
>Centre
Move
Left
Move
Right
Logging
Log
Type
Log
Trigger
Timer
Setup
Logging
Control
Auto
Measure
On/Off
Mode
Measure
Type
Set Chn
B/W