User Manual
Page 52 
Amplitude Markers 
Number of Markers: 
One, Two, or None 
Marker Resolution: 
0.1 dB 
Amplitude Readout: 
Display of absolute and difference amplitudes for both markers in 
graticule units (dBm or dBuV) or linear units (mV or uW). 
Amplitude Limits (only with Option U01) 
Limit Types: 
Limit Lines from numeric values, or Limit Patterns from Files. Pattern 
files are created using PSA-Manager software. 
Number of Limits: 
One or two limits on screen. Up to 999 patterns can be stored 
Limits Comparator: 
Conditions of Above, Below, Inside or Outside creating actions of 
beep, stop sweep, log sweep and pulse out  
Amplitude Compensation (only with Option U01) 
Offset: 
Amplitude can be offset by up to +/-50dB to compensate for external 
attenuation or gain. 
75  Compensation 
Compensation can be made for inputs from a 75  source impedance. 
Compensation 
Tables: 
Linearly interpolated tables of up to 100 frequency/amplitude points 
can be used. Up to 999 tables can be stored. Table files are created 
using PSA-Manager software. 
12.1.3  Traces 
Graticule: 
8.5 x 10 divisions light grey graticule. Selectable as normal, dim, or off. 
Displayed Points: 
271 points per sweep (peak detected). 
Live Trace: 
Dot-joined trace from current or held sweep. Selectable on or off. 
View Trace 
Additional trace created from the live trace. Selectable on or off. 
Reference Trace: 
Trace recalled from a file. Selectable on or off. 
Trace Modes: 
Normal, Peak Hold, or Average (2 to 48 sweeps) 
12.1.4  Sweep 
Sweep Method: 
Peak detection for 270 points per sweep. The amplitude of the peak 
level found within each sub-span is stored (sub-span = span/270). 
Sweep Time: 
Set automatically by Span and RBW. 
Typically 150ms + 2.3ms/MHz of span for RBW = 1MHz  
Typically 150ms + 7ms/MHz of span for RBW = 280kHz 
Typically 180ms + 77ms/MHz of span for RBW = 15kHz 
Sweep Modes: 
Normal (repeat) or Single 
Sweep Trigger 
(Only with Option U01), External or Limits Comparator 
12.1.5  Data Logging (only with Option U01) 
Log Files 
Files of up to 9,999 data entries can be created. Up to 999 files can be 
stored. Files are read using PSA-Manager Software. 
Logged Data: 
Log files can be created from Peak Value, Centre Frequency Level, 
Complete Sweep, or Screen Image 
Log Trigger: 
Log entry can be triggered by Key Press, Continuous (every sweep), 
External Input, Internal Timer, or Limits Comparator 
Timer Interval 
Adjustable between 2 seconds and 100 minutes per entry 
Time Stamp 
Each entry is time stamped from the real-time clock 