User Manual
Page 49 
10  EMC 
Spectrum Analyzer – powered by internal battery or Adaptor/Charger 
This instrument has been designed to meet the requirements of the EMC Directive 89/336/EEC. 
Compliance was demonstrated by meeting the test limits of the following standards: 
Emissions 
EN61326-1 (2006) EMC product standard for Electrical Equipment for Measurement, Control and 
Laboratory Use.  Test limits used were: 
a)  Radiated: Class B 
Note: Limits were met without a test object connected; emissions which exceed the levels required 
by this standard may occur when the instrument is connected to a test object. 
Immunity 
EN61326-1 (2006) EMC product standard for Electrical Equipment for Measurement, Control and 
Laboratory Use. 
Test methods, limits and performance achieved were: 
a)  EN61000-4-2 (2009) Electrostatic Discharge : 4kV air, 4kV contact, Performance A (A). 
b)  EN61000-4-3 (2006) Electromagnetic Field, 3V/m, 80% AM at 1kHz, Performance B (A). 
Note: The PSA1302 & PSA2702 are sensitive measuring instruments and, if 
subjected to a sufficiently large RF field, the interfering signal may show a few dB 
above the noise floor even with the SMA input capped (Performance B). In all other 
respects the instrument will operate correctly (Performance A) in fields up to 3V/m. 
Adaptor/Charger 
This AC adaptor/charger has been designed to meet the requirements of the EMC Directive 
89/336/EEC. Compliance was demonstrated by meeting the test limits of the following standards: 
Emissions 
EN55022, radiated and conducted Class B. 
Immunity 
EN55024:1998 + A2:2003. Test methods, limits and performance achieved were: 
a)  EN61000-4-2 (2001) Electrostatic Discharge : 4kV air, 4kV contact, Performance A (B). 
b)  EN61000-4-3 (2002) Electromagnetic Field: 3V/m, 80% AM at 1kHz, Performance A (A). 
c)  EN61000-4-11 (2004) Voltage Interrupt: ½ cycle 0%: Performance A (B); 1 cycle 0% 
Performance B (B); 25 cycles, 70% and 250 cycles, 0%: Performance B (C). 
d)  EN61000-4-4 (2004) Fast Transient: 1kV peak (AC line), Performance A (B)   
e)  EN61000-4-5 (2001) Surge: 1kV (line to line), 2kV (line to ground), Performance A (B). 
f)  EN61000-4-6 (2003) Conducted RF: 3V, 80% AM at 1kHz (AC line & DC Output) 
Performance A (A). 
Performance Definitions 
The definitions of performance criteria are: 
Performance criterion A: ‘During test normal performance within the specification limits.’ 
Performance criterion B: ‘During test, temporary degradation, or loss of function or performance 
which is self-recovering’. 
Performance criterion C: ‘During test, temporary degradation, or loss of function or performance 
which requires operator intervention or system reset occurs.’ 