4.7 Setup/Functions
4.7.1 Setup/Functions > Logging
Logging functions are only available when Option U01 is installed.
Log Type creates a pop-up menu that selects the type of data to be logged. Options
are Centre Level, Peak Level, Full Trace, or Screen Image.
Log Trigger creates a pop-up menu that selects the trigger for logging entries.
Options are Timer, Manual (using Run/Stop key), External trigger input (+ve or -ve
level change), Limits Event or Continuous (log after every sweep).
Timer Setup sets the internal timer interval between 1 second and 99 minutes. Note
that the actual logging interval will depend upon the sweep time which may be much
longer.
Logging Control opens a control menu that puts the instrument into logging mode.
No changes to the instrument set-up can be made once logging is commenced.
All logging conditions must be set prior to using the logging control screen.
The Logging control screen provides an overview of the logging set-up and has keys as follows:
Change Default the auto-incrementing default name for log files can be changed to any five
character name.
File Utilities opens a file list screen for log files enabling renaming and deletion.
Change Name enables a specific name for the new log file (rather than default name) to be
entered.
Enable Logging opens the new log file and enables logging entries to the file (a warning screen
is displayed requiring confirmation).
Pause/Resume logging entries can be paused, but no changes to the set-up can be made.
Close File disables logging and closes the file. Once closed, no further entries can be made.
Cancel/Exit exits the control screen if no log file is open. If a log file is open, it pauses logging
and provides options to resume or to close and exit.
4.7.2 Setup/Functions > Setups
Enables the complete set-up of the instrument to be stored to a file and recalled
later.
Presets operates as a shortcut to the Presets menu. See section 2.8.3.
Store Setup creates a new control screen from which the setup can be saved.
Save saves the file under a default name that auto-increments from 001 to 999.
Change Name brings up an alpha-numeric keypad from which a new file name of up
to eight characters can be entered.
Change Default enables the default file name used to be changed. Up to five characters can be
used.
File Utilities brings up a file window from which all existing files can be seen, and allows them
be deleted or renamed.
Recall Setup brings up a file window from which the chosen file can be selected and
recalled. Files are listed by date.
Find A-Z lists only files starting with that character.
Latest Files lists up to 20 recently stored or recalled files.
Load Defaults sets the instrument back to the settings that existed when the
instrument was manufactured. All settings including screen brightness and off-timer
are reset. However, no files are deleted.