Instructions

7
CURRENT MONITOR OUTPUT
Output Terminals:
BNC (chassis ground) on front panel or terminal block on rear panel.
Output Impedance:
600 nominal, for >1M load (e.g. oscilloscope).
Scaling:
250mV per Amp (4 Volts full scale).
Accuracy:
± 0·5% ± 5mV.
Isolation:
CATII (300V) to load negative.
Bandwidth limit (-3dB):
40kHz.
REMOTE CONTROL
Digital Remote Interfaces
The unit provides LAN, USB, GPIB and RS232 interfaces for full remote control.
LAN:
Ethernet 100/10base-T connection with auto cross-over detection.
1.4 LXI Core 2011 compliant.
USB:
Standard USB 2.0 connection. Operates as virtual COM port.
GPIB: Conforming to IEEE488.1 and IEEE488.2.
Capabilities: SH1, AH1, T6, L4, SR1, RL2, PP1, DC1, DT0, C0, E2.
RS232:
Standard 9-pin D connection. Baud rate: 9600.
External Control Input Characteristics
Connection: Terminal block on rear panel. Lo terminal input grounded to chassis
internally.
10k. Input protected against excess input voltages up to 50V.
CATII (300V) to load negative.
External Analogue Voltage Control
Operating Mode:
The applied voltage sets the operating level within the range.
Scaling:
4 Volts full scale (250mV per Amp).
Accuracy:
± 2% ± accuracy of range.
Common mode rejection:
Typically better than 76dB.
External Logic Level (TTL) Control
Operating Mode:
The applied signal selects between Level A and Level B settings.
Threshold:
+ 1·5V nominal. A logic high selects Level B.
Remote Disable Input
Connection: Terminal block on rear panel.
Input to the LED of an opto-isolator through 1k resistor.
Threshold:
Apply >+3V to disable the load input. Max. Voltage 12V.