User Manual
Specification
Specifications apply for 18ºC − 28ºC ambient after 30 minute warm-up.
Functions
Parameters Measured: R, L, C, D & Q.
Measurement Modes: Series or parallel equivalent circuit.
Measurement Functions: Fully autoranging including selection between L, C and R. The Zero C
function nulls out up to 100pF of stray capacitance in the test fixture.
Measurement Frequency: User selectable to be 100Hz, 1kHz or 10kHz; frequency accuracy
± 0.01%. 120Hz instead of 100Hz by factory option for 60Hz operation.
Measurement Ranges
and Resolution:
Parameter
Range
R
0.1mΩ − 990MΩ
L
0.001µH − 9900H
C
0.001pF − 99000µF
D
0.001 − 999
Q
0.001 − 999
Measurement Accuracy:
100/120Hz 1kHz 10kHz
R (Q<0·1) 0.1% ± 1 digit
0.5% ± 1 digit
2% ± 1 digit
2Ω − 1MΩ
0.4Ω − 5MΩ
0.1Ω − 20MΩ
2Ω − 500kΩ
0.4Ω − 2MΩ
0.1Ω − 10MΩ
2Ω − 50kΩ
0.4Ω − 200kΩ
0.1Ω – 500kΩ
L (Q>10) 0.1% ± 1 digit
0.5% ± 1 digit
2% ± 1 digit
4mH – 500H
800µH – 2500H
200µH − 9900H
400µH − 50H
80µH – 250H
20µH − 1000H
40µH − 5H
8µH – 25H
2µH − 100H
C (D<0.1) 0.1% ± 1 digit
0.5% ± 1 digit
2% ± 1 digit
10nF − 1000µF
2nF − 5000µF
500pF − 20000µF
1nF − 100µF
200pF − 500µF
50pF − 2000µF
100pF − 10µF
20pF − 50µF
5pF − 200µF
Q & D 0.25% ± 1 digit
0.25 − 4.0
for C = 40nF – 100µF
or L = 10mH – 50H
0.25 − 4.0
for C = 10nF – 10µF
or L = 1mH – 2.5H
0.25 − 4.0
for C = 1nF – 1µF
or L = 100µH – 250mH
Capacitance accuracies apply after null.
Measurement Update Rate: 2.5 readings per second.
Limits Comparator (Sort Mode)
Type: Comparison with multiple limits set up from the keyboard or PC via RS232 interface.
Binning: Up to 8 Pass bins for the major parameter, plus minor parameter Fail and general Fail bins.
3