19 Specifications
ACCURACY
Accuracies apply for 1 year 19°C to 25°C. Temperature coefficient outside these limits is <0.1 x
quoted range accuracy per °C.
DC Volts
Range
Accuracy
Resolution
Input Impedance
100mV
0.02% ± 3 dig. (after null)
1uV
10.11M
Ω
1000mV
0.02% ± 3 dig.
10uV
10.11M
Ω
10V
0.02% ± 3 dig.
100uV
10.0M
Ω
100V
0.02% ± 3 dig.
1mV
10.1M
Ω
1000V
0.02% ± 3 dig.
10mV
10.11M
Ω
Typical Input
Impedance:
10MΩ//<1000pF, except for Vdc plus Vac measurement when the
1MΩ ac attenuator is in parallel with the 10MΩ dc attenuator.
Maximum Input:
1kV DC or AC peak, any range.
NMR:
>60dB at 50/60Hz.
1kΩ Unbalanced CMR:
>90dB at DC/50Hz/60Hz.
Recommended maximum peak AC plus DC voltage while measuring DC voltage
These values only apply when the DC component does not exceed the range maximum and the
analogue 50/60Hz rejection filter is enabled.
AC Volts (True RMS)
Range Accuracy Resolution
45Hz - 10kHz 10kHz - 30kHz 30kHz - 50kHz
100mV 0.2% ± 150 dig 1.5% ± 200 dig. - 1µV
1000mV 0.2% ± 100 dig 0.5% ± 100 dig 2% ± 200 dig 10µV
10V 0.2% ± 100 dig 0.5% ± 100 dig 2% ± 200 dig 100uV
100V 0.2% ± 100 dig 0.5% ± 100 dig 2% ± 200 dig 1mV
750V 0.2% ± 100 dig 0.5% ± 100 dig 2% ± 200 dig 10mV
Accuracy specifications apply for readings between 10,000 and 120,000 counts. Maximum crest
factor = 3 at nominal range maximum.
Input Impedance:
1MΩ//<1000pF
Maximum Input:
750V rms, 1000V peak; any range.
1kΩ Unbalanced CMR:
>60dB at DC/50Hz/60Hz
(AC + DC) Voltage Accuracy
Total measurement error will not exceed the sum of the separate ac and dc accuracy specifications
plus 10 display counts.
Range
AC
pk
+DC
Range
AC
pk
+DC
The table shows the maximum peak AC
(45Hz-50kHz) plus DC voltage that can be
applied without affecting the accuracy of the
DC measurement due to overload clamping.
100mV, 1000mV
1.8V
100V
180V
10V
18V
1000V
1000V