Instructions
Page 50 
 Resistance  
2-wire or 4-wire measurement. 
Range 
Accuracy 
Resolution 
100Ω 
0.05% ± 8 dig. 
1mΩ 
1000Ω 
0.05% ± 5 dig. 
10mΩ 
10kΩ 
0.05% ± 5 dig. 
100mΩ 
100kΩ 
0.05% ± 5 dig. 
1Ω 
1000kΩ 
0.05% ± 5 dig. 
10Ω 
10MΩ 
0.3% ± 2 dig. 
100Ω 
Maximum Input: 
1000V DC or AC rms, any range. 
Maximum Open Circuit Voltage: 
3.5V 
DC Current  
Range 
Accuracy 
Resolution 
10mA 
0.05% ± 5 dig. 
100nA 
100mA 
0.05% ± 5 dig. 
1uA 
1A 
0.2% ± 5 dig. 
10uA 
10A 
0.2% ± 5 dig to 5A 
100uA 
10A 
0.5% ± 10 dig to 10A 
100uA 
Maximum Input: 
mA ranges - 1.6A DC or AC rms, 1000V, fuse protected. 
10A range - 10A DC or AC rms, 1000V, fuse protected. 
Typical Voltage Burden: 
10mA range - <100mV 
10A, 1A and 100mA  ranges - <600mV 
AC Current (True RMS)  
Range 
Accuracy (45Hz - 10kHz) 
Resolution 
10mA 
0.35% ± 20 dig. 
100nA 
100mA 
0.35% ± 20 dig. 
1uA 
1A 
0.5% ± 20 dig.  
10uA 
10A 
0.5% ± 20 dig to 5A 
100uA 
10A 
1% ± 20 dig. to 10A 
100uA 
Accuracy specifications apply for readings between 10,000 and 120,000 counts. Additional error at 
crest factory = 3 typically 0.5%. 
Maximum Input: 
mA ranges - 1.6A DC or AC rms, 1000V, fuse protected. 
10A range - 10A DC or AC rms, 1000V, fuse protected. 
Typical Voltage Burden: 
10mA range - <100mV 
10A, 1A and 100mA ranges - <600mV 
(AC + DC) Current Accuracy 
Total measurement error will not exceed the sum of the separate ac and dc accuracy specifications 
plus 10 display counts. 