Instructions
Page 45 
16.2.7  Common Commands  
*IDN? 
Returns the instrument identification.  
The response is in the form <NAME>, <model>, <serial>, <version><RMT> 
where <NAME> is the manufacturer's name, <model> is the instrument 
type, <serial> is the interface serial number and <version> is the revision 
level of the firmware installed. 
*RST 
Resets the functional parameters of the instrument to the default settings as 
listed in the Factory Default Settings section. 
Does not affect any remote interface settings.  
*OPC 
Sets the Operation Complete bit (bit 0) in the Standard Event Status 
Register. This will happen immediately the command is executed because 
of the sequential nature of all operations. 
*OPC? 
Query Operation Complete status.  
The response is always 1<RMT> and is available immediately the command 
is executed because all commands are sequential.  
*WAI 
Wait for Operation Complete true.  
This command does nothing because all operations are sequential. 
*TST? 
The DMM has no self-test capability and the response is always 0
<RMT>
. 
*TRG 
The DMM has no trigger capability. The command is ignored in this 
instrument. 
16.2.8  Status Commands 
*CLS 
Clear Status. Clears all status indications, including the Status Byte.  
Does not clear any Enable Registers. 
ITR? 
Query the Input Trip Register. The response format is 
<NR1><RMT>
.  
Clears any bits that no longer apply. 
ITE 
<NRF>
Set the Input Trip Enable Register to 
<NRF>
ITE? 
Returns the value in the value in the Input Status Enable Register.  
The response is: <NR1><RMT>. 
EER? 
Query and clear Execution Error Register. The response format is 
<NR1><RMT>. 
QER? 
Query and clear Query Error Register. The response format is 
<NR1><RMT>
. 
*STB? 
Report the value of the Status Byte. The response is: 
<NR1><RMT>
. 
Because there is no output queue, MAV can only be read by a GPIB serial 
poll, not by this query, as any previous message must have already been 
sent. 
*SRE 
<NRF>
Sets the Service Request Enable Register to 
<NRF>
*SRE? 
Report the value in the Service Request Enable Register.  
The response is <NR1><RMT>. 
*PRE 
<NRF>
Set the Parallel Poll Enable Register to the value 
<NRF>
. 
*PRE? 
Report the value in the Parallel Poll Enable Register.  
The response is <NR1><RMT>. 
*IST? 
Returns the state of the ist local message as defined by IEEE Std. 488.2.  
The response is 0<RMT> if the local message is false, or 1<RMT> if true. 