Instructions
Page 20 
Main 
Secondary 
Measurement Time 
(4SPS) 
Measurement Time 
(20SPS) 
Any Dual Function 
None 
.25s 
.05s 
Vac, dB, Iac 
Hz* 
.25s 
.05s 
Hz* 
Vac, Iac 
.25s 
.05s 
Vac, dB, Iac 
Vdc, Idc 
.5s 
.1s 
Vdc, Idc 
Vac, Iac 
.5s 
.1s 
Vac, dB 
Iac 
6s 
3s 
Iac 
Vac 
6s 
3s 
Vdc + Vac 
Vdc, Vac, Hz 
.5s 
.1s 
Idc + Iac 
Idc, Iac, Hz 
.5s 
.1s 
Measurement Time 
(4SPS) 
Measurement Time 
(20SPS) 
*Any Hz dual measurement 
.25s 
.125s 
Frequency measurements have a fixed measurement time (gate time) of 0.25s or 0.125s 
depending on the sample speed. This fixed measurement time only applies to the frequency 
measurement, not the other dual measurement. For example if the sample speed is set to fast 
(20SPS), the main measurement is Hz and the secondary measurement is VAC, then VAC will 
be updated every 0.05s but the frequency measurement will only update every .125s.  
9  Triggered Operation 
A screw-less connector on the rear panel provides both an input and output signal. 
9.1  Trigger In Signal 
The trigger input (- I/P+) connects to the LED of an opto-isolator through a 1kΩ resistor. 
Apply >+3V to set trigger input True; maximum safe applied input is +10V. 
The trigger input can be set to perform either of two actions, as set from the Utilities menu – see 
section 12.4
. 
1.  Hold measurement – the measurement in progress at the point that the input is set True 
is frozen on the display. 
2.  Log measurement – with the Logging function operating, the measurement in progress 
at the point that the input is set True will be logged. 
9.2  Trigger Out Signal (Measurement Complete) 
The trigger out signal (- O/P+) goes True each time an input trigger action is completed. 
The trigger out signal is an opto-isolated open collector output with Low = True. The output 
conducts for 250ms on completion of an input trigger action. 
The trigger output requires an external pull-up (e.g. 4.7kΩ to 5V) and sinks typically 2mA 
maximum when True; maximum safe off-state voltage is +10V. 