Data Sheet

Technical Specifications - 1908 and 1908P
MEASUREMENT PARAMETERS & ACCURACY
Accuracies apply for 1 year 19°C to 25°C. Temperature coefficient outside these limits
is <0.1 x quoted range accuracy per °C.
Scale Length
The scale length is 120,000 counts (unless otherwise stated) giving a maximum
measured value for each range of 119999. Thus for the 100V range the maximum
measured value is 119.999 volts
Accuracies are quoted in terms of a percentage of the measured value plus a number
of least significant digits.
Reading Rate
Varies with measurement parameter. Selectable as 4 readings/second or 20
readings/second for most parameters.
DC Voltage
Range Accuracy Resolution Input Impedance
100mV 0.02% ± 3 digits* 1µV 10.11M
1000mV 0.02% ± 3 digits 10µV 10.11M
10V 0.02% ± 3 digits 100µV 10.0M
100V 0.02% ± 3 digits 1mV 10.1M
1000V 0.02% ± 3 digits 10mV 10.11M
* after null
Typical Input 10MW//<1000pF, except for Vdc plus Vac measurement when the
Impedance: 1MW AC attenuator is in parallel with the 10MW DC attenuator.
Maximum Input: 1kV DC or AC peak, any range.
NMR: >60dB at 50/60Hz.
CMR: 1kW unbalanced CMR is >90dB at DC/50Hz/60Hz.
Recommended max.peak AC + DC voltage while measuring DC voltage
Range 100mV 1000mV 10V 100V 1000V
Max. AC 1.8V 1.8V 18V 180V 1000V
The table shows the maximum peak AC (45Hz-50Hz) plus DC voltage that can be
applied without affecting the accuracy of the DC measurement due to overload
clamping. These values only apply when the DC component does not exceed the range
maximum and the analogue 50/60Hz rejection filter is enabled.
AC Voltage (True RMS)
Range Accuracy Resolution
45Hz - 10kHz 10kHz - 30kHz 30kHz - 50kHz
100mV 0.2% ± 150 dig. 1.5% ± 200 dig. - 1µV
1000mV 0.2% ± 100 dig. 0.5% ± 100 dig. 2% ± 200 dig. 10µV
10V 0.2% ± 100 dig. 0.5% ± 100 dig. 2% ± 200 dig. 100µV
100V 0.2% ± 100 dig. 0.5% ± 100 dig. 2% ± 200 dig. 1mV
750V 0.2% ± 100 dig. 0.5% ± 100 dig. 2% ± 200 dig. 10mV
Accuracy specifications apply for readings between 10,000 and 120,000 counts.
Maximum crest factor = 3 at nominal range maximum.
Input Impedance: 1MW//<1000pF
Maximum Input: 750V rms, 1000V peak; any range.
CMR: 1kW Unbalanced CMR is >60dB at DC/50Hz/60Hz
AC + DC Voltage
Total measurement error will not exceed the sum of the separate AC and DC accuracy
specifications plus 10 digits.
Resistance
2-wire or 4-wire measurement
Range Accuracy Resolution
100W 0.05% ± 8 digits 1mW
1000W 0.05% ± 5 digits 10mW
10kW 0.05% ± 5 digits 100mW
100kW 0.05% ± 5 digits 1W
1000kW 0.3% ± 2 digits 10W
10MW 0.3% ± 2 digits 100W
Maximum Input: 1000V DC or AC rms, any range.
Max. O/C Voltage: 3.5V
DC Current
Range Accuracy Resolution
10mA 0.05% ± 5 digits 100nA
100mA 0.05% ± 5 digits 1µA
1000mA 0.2% ± 5 digits 10µA
10A (<5A) 0.2% ± 5 digits 100µA
10A (>5A) 0.5% ± 10 digits 100µA
Maximum Input: mA ranges - 1.6A DC or AC rms, 1000V, fuse protected.
10A range - 10A DC or AC rms, 1000V, fuse protected.
Voltage Burden: 10mA range - <100mV; 100mA, 1A, 10A ranges - <600mV
AC Current (True RMS)
Range Accuracy (45Hz to 10kHz) Resolution
10mA 0.35% ± 20 digits 100nA
100mA 0.35% ± 20 digits 1µA
1000mA 0.5% ± 20 digits 10µA
10A (<5A) 0.5% ± 20 digits 100µA
10A (>5A) 1% ± 20 digits 100µA
Accuracy specifications apply for readings between 10,000 and 120,000 counts.
Maximum crest factor = 3 is typically 0.5%.
Maximum Input: mA ranges - 1.6A DC or AC rms, 1000V, fuse protected.
10A range - 10A DC or AC rms, 1000V, fuse protected.
Voltage Burden: 10mA range - <100mV; 100mA, 1A, 10A ranges - <600mV
AC + DC Current
Total measurement error will not exceed the sum of the separate AC and DC accuracy
specifications plus 10 digits.
Frequency
Range Accuracy Resolution
100Hz 0.01% ± 1 digit 0.01Hz
1000Hz 0.01% ± 1 digit 0.1Hz
10kHz 0.01% ± 1 digit 1Hz
100kHz 0.01% ± 1 digit 10Hz
Scale Length: 12,000 counts
Frequency Range: 10Hz to >100kHz
Input sensitivity: Better than 30mVrms (100mV range);
better than 10% of range for all other Vac and Iac ranges.
Capacitance
Range Accuracy Resolution
10nF 2% ± 5 digits 10pF
100nF 2% ± 5 digits 100pF
1µF 2% ± 5 digits 1nF
10µF 2% ± 5 digits 10nF
100µF 5% ± 5 digits 100nF
Scale Length: 1200 counts
Temperature
Linearised measurement for PT100 or PT1000 RTD probes, 2-wire or 4-wire
measurement, display in
o
C or
o
F.
RTD Range Accuracy Resolution
PT100 -50 to 400
o
C 0.05% ± 0.5
o
C 0.1
o
C
PT1000 -50 to 400
o
C 0.05% ± 0.5
o
C 0.1
o
C
N. B. Accuracy specifications apply to multimeter only, they do not account for additional probe accuracy
errors.
Continuity and Diode Test
Continuity: 100W range selected; audible tone sounds for impedance <10W.
Diode Test: Test current approximately 1mA; displays voltages up to 1.2V.
Max. O/C Voltage: 3.5V
Maximum Input: 1000V DC or AC rms, any range.