User's Guide

30
Electrostatic
discharge
(ESD)
IEC 60601-4-2
±6KV Contact
±8KV Air
±6KV Contact
±8KV Air
Floors should be wood, concrete
or ceramic tile. If floors are
covered with synthetic material,
the relative humidity should be
at least 30 %.
Electrical fast
transient burst
IEC 61000-4-4
±2KV Power cord
±1KV input/output
±2KV Power Cord
±1KV input/output
Mains power quality should be
that of a typical commercial or
hospital environment.
Surge
IEC 61000-4-5
±1KV Line to GND
±2KV Line to GND
±1KVLine to GND
±2KV Line to GND
Mains power quality should be
that of a typical commercial or
hospital environment.
Voltage dips,
short
interruptions
and
voltage
variations
on power
supply
input lines
IEC 61000-4-11
<5% UT for 0.5
weeks (>95% dip in
UT) 40% UT for 5
weeks (60% dip in
UT) 70% UT for 25
weeks (30% dip in
UT) <5% UT for 5s
(>95% dip in UT)
<5% UT for 0.5
weeks (>95% dip in
UT) 40% UT for 5
weeks (60% dip in
UT) 70% UT for 25
weeks (30% dip in
UT) <5% UT for 5s
(>95% dip in UT)
Mains power quality should be
thatof a typical commercial or
hospitalenvironment. If the user
of the devicerequires continued
operation duringpower mains
interruptions, it isrecommended
that the device bepowered from
an uninterruptible powersupply
or a battery.
Power
frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3A/m
3A/m
The power frequency magnetic
fieldshould have the
characteristics of
powerfrequency magnetic field
level at atypical place in a typical
commercial andhospital
environment
IEC 60601-1-2: Table 204
Guidance and manufacturer's declaration electromagnetic immunity
The device is intended for use in the electromagnetic environment specified below. The
customer or the user of the device should ensure that it is used in such an environment.
Immunity Test
IEC 60601 Test Level
Compliance Level
Electromagnetic environment -
guidance
Conducted RF
IEC 61000-4-6
3VVrms
150kHz~80MHz
10VEngineering
medical
frequency
band
3VVrms
10V
(Engineering
medical
frequency
band
Portable and mobile RF
communications equipment
should be used no closer to any
part of the device,
including cables, than the
recommended
separation distance calculated
from the equation
applicable to the frequency of