User's Guide
12
When the sensor initialization is complete, the PDA will begin to display the
current blood sugar value.
If the sensor has previously been implanted and connected to the signal
transmitter for a period of time, after disconnecting and reconnecting with the
signal transmitter, click the "Cancel" button when prompted by the portable
monitor to "detect the new sensor and confirm".
At this point, the portable monitor skips initialization and starts displaying the
current blood sugar value directly.
<Note>Do not cancel the initialization process when using the new sensor.
Skipping the initialization of the new sensor may cause the system to display
incorrect glucose concentration values.</
注意>
Sensor Calibration