User's Guide

24
Wash and dry your hands before obtaining a blood sample.
Make sure you closely follow the instructions in this manual.
Blood oxygen and red blood cell count vary from person to person, and
even within the same person. The Exactive Vital glucose meter tests blood
glucose concentrations for the widest range of people possible. If the
users blood indexes fall within the middle of the range, the result will
be ideal. Otherwise, there will be some small deviations. (The deviations
should be within the range allowed by local government.)
4.Quality Control Tests
Control solution is a glucose solution of known concentration that
is used to confirm that your PDA blood glucose meter and test strips are
working properly.Control solution 1 is usually used first and control
solution 2 is used when a second level test is required. MicroTech offers
a separately available control solution package that contains both
Control Solution 1 and Control Solution 2. Always use a control solution
for quality control testing to ensure you get accurate blood glucose
levels.
You should perform the control solution test below if you suspect that
the meter or test strips are not working properly or when you suspect that
your test results are inaccurate, or inconsistent with how you feel
When the PDA display is on, just insert a test strip into the PDA’s
test strip portand the blood glucose test screen will appear.Check
the Control Solution Test check box at the bottom.The animation
will be replaced with an animation of a control solution drop.Shake
the control solution bottle, gently squeeze out the control solution,
discard the first drop, and drop the second drop onto a clean
nonabsorbent surface. Now touch the second drop to the sample area
of the test strip. Do not let the bottle come into contact with the
strip itself.
When enough control solution has been applied, the screen will count
down 5 seconds and then display the test result. The result is
displayed in the top half of the screen. If the result falls within