Agilent Technologies X/P/K281C Adapters Including Options 006, 012, 013 Operating and Service Manual Agilent Part Number: 00281-90043 Printed in USA Print Date: April 2002 Supersedes: June 2001
Notice The information contained in this document is subject to change without notice. Agilent Technologies makes no warranty of any kind with regard to this material, including, but not limited to, the implied warranties of merchantability and fitness for a particular purpose. Agilent Technologies shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing, performance, or use of this material.
Warranty Certification Agilent Technologies certifies that this product met its published specifications at the time of shipment from the factory. Agilent Technologies further certifies that its calibration measurements are traceable to the United States National Institute of Standards and Technology (NIST, formerly NBS), to the extent allowed by the Institute’s calibration facility, and to the calibration facilities of other International Standards Organization members.
Assistance Product maintenance agreements and other customer assistance agreements are available for Agilent Technologies products. For assistance, call your local Agilent Technologies Sales and Service Office (refer to “Service and Support” on page iv). Service and Support By internet, phone, or fax, get assistance with all your test and measurement needs. Online assistance: www.agilent.
Safety and Regulatory Information Review this product and related documentation to familiarize yourself with safety markings and instructions before you operate the instrument. This product has been designed and tested in accordance with international standards. WARNING The WARNING notice denotes a hazard. It calls attention to a procedure, practice, or the like, that, if not correctly performed or adhered to, could result in personal injury.
This ISM devise complies with Canadian ICES-001. Cet appareil ISM est conforme a la norme NMB du Canada This symbol indicates that the power line switch is ON. This symbol indicates that the power line switch is OFF or in STANDBY position. Safety Earth Ground This is a Safety Class I product (provided with a protective earthing terminal). An uninterruptible safety earth ground must be provided from the main power source to the product input wiring terminals, power cord, or supplied power cord set.
General Information General Information Description The X281C, P281C, and K281C adapters provide a convenient means of coupling between waveguide and coaxial systems. Power can be transmitted in either direction, and each adapter covers the full frequency range of its waveguide size. A step-like internal structure transforms the waveguide impedance to the 50 Ω impedance of the coaxial line. Option 006 Table 1 Option 006 adds two alignment holes to the waveguide flange.
General Information Option 012 Option 012 for the X281C offers a Type-N (m) connector to replace the standard 7-mm connector. Option 013 Option 013 for the X281C offers a Type-N (f) connector to replace the standard 7-mm connector. Option 013 is not offered for the K281C and the P281C. Instruments Covered by Manual The adapters covered by this manual have a two part serial number. The first four digits and letter constitute the serial number prefix.
Specifications Specifications Specifications shown in Table 2 are the performance standards against which the adapter may be tested. Table 2 Specifications X281C P281C K281C 8.2 to 12.4 12.4 to 18 18 to 26.5 < 1.05 < 1.06 < 1.07 (Typical SWR) (< 1.03) (< 1.04) (< 1.05) Operating temperature 0 to +55 °C 0 to +55 °C 0 to +55 °C Frequency range (GHz) SWR 1 1. Specifications in this table are measured with no gap between the full diameters of the male and female center conductors.
Specifications Supplemental characteristics listed in Table 3 are non-warranted performance parameters that are included as additional information for the user. Table 3 Supplemental Characteristics X281C P281C K281C Typical insertion loss 0.08 dB 0.10 dB 0.12 dB Maximum peak power1 200 W 200 W 100 W 25.40 x 12.70 (mm) 17.83 x 9.93 (mm) 12.70 x 6.35 (mm) 1.00 x 0.50 (in) 0.70 x 0.39 (in) 0.50 x 0.
Operation Operation Before connecting the adapter, note the precautions below. CAUTION Operating Environment Operating Procedures • Exceeding the energy and power levels shown in Table 1 may result in damage to the adapter or associated equipment. • Care should be taken to protect the face of the flange from any damage that would prevent close surface-to-surface contact. Any burring, denting, or scratching may increase RF leakage and the reflection coefficient of the waveguide connection.
Service Information Service Information Replacing the Center Conductor Contact Not all parts on the adapters are replaceable, but all of the adapters have a replaceable center conductor contact. When you replace the center conductor contact, observe the following precautions: • Before installing the new center conductor contact, apply Loctite sealant #222 to the threads of the center conductor. • Do not use excessive force when tightening center conductor parts.
Replaceable Parts Replaceable Parts Replaceable parts for P281C adapter are shown below. Refer to Table 4, Table 5, and Table 6 for identification of parts and corresponding part numbers for all models. For ordering information, contact the nearest Agilent office listed in “Service and Support” on page iv. Figure 3 Replaceable Parts NOTE Table 1 on page 1 gives dimensional tolerances for Option 006 that can be expected to yield satisfactory results.
Replaceable Parts Table 4 X281C Replaceable Parts Item No. Description Part Number 1 Connector body 1250-1466 2 Spacer 00281-20027 (0.05 mm) or 00281-20028 (0.075 mm) or 00281-20049 (0.025 mm) 3 Center conductor contact G-slotted 85050-20001 Collet holder 85130-20002 Shim 5020-8540 (0.013 mm) or 5020-8541 (0.
Replaceable Parts Table 4 X281C Replaceable Parts (Continued) Item No. Description Part Number X281C Option 013 Table 5 1 Connector body 1250-0914 2 Spacer 00281-20027 (0.05 mm) or 00281-20028 (0.075 mm) or 00281-20049 (0.025 mm) 3 Center conductor contact 5180-0854 4 Shim 5020-8540 (0.013 mm) or 5020-8541 (0.
Replaceable Parts Table 6 K281C Replaceable Parts Item No. Description Part Number 1 Connector body 1250-1507 2 Spacer 5021-9679 (0.013 mm) or 5021-9680 (0.025 mm) or 5021-9681 (0.051 mm) 3 Center conductor contact 00281-20043 6 Spacer 000281-20045 7 Washer 3050-0261 8 Nut 0608-0003 9 Blank label Not replaceable 10 Flange cap 5040-0357 Needed But Not Supplied 5/64 inch hexagonal nut driver K218C Option 012 1 Connector body 1250-1509 2 Spacer 5021-9679 (0.
Connector Types Connector Types Pin depth is the distance that the center conductor mating plane differs from being flush with the outer conductor mating plane. The pin depth of a connector can be in one of two states, protruding or recessed. Figure 4, Figure 5, and Figure 6 show a visual representation of proper pin depth for connector types for standard and option models. Table 7 Connector Types for Standard and Options X281C P281C K281C Standard 7-mm 7-mm APC-3.
Connector Types 3.5-mm Male Connector 3.5-mm Female Connector See Detail "A" See Detail "B" Center Conductor Mating Plane Center Conductor Mating Plane Detail "A" Detail "B" Outer Conductor Mating Plane Outer Conductor Mating Plane Pin Depth Figure 5 Pin Depth 3.